High Intensity Focused Ultrasound Lesion Detection Using Adaptive Compressive Sensing Based on Empirical Mode Decomposition
DOI: 10.4103/jmss.JMSS_17_18
Abstract
Background: The main goal of ultrasound therapy is to have clinical effects in the tissue without damage to the intervening and surrounding tissues. Treatments have been developed for both in vitro and in clinical applications. HIFU therapy is one of these. Non-invasive surgeries, such as HIFU, have been developed to treat tumors or to stop bleeding. In this approach, an adequate imaging method for monitoring and controlling the treatment is required. Methods: In this paper, an adaptive compressive sensing representation of ultrasound RF echo signals is presented based on empirical mode decomposition (EMD). According to the different numbers of intrinsic mode functions (IMFs) produced by the EMD, the ultrasound signals is adaptively compressive sampled in the source and then adaptively reconstructed in the receiver domains. In this paper, a new application of compressive sensing based on EMD (CSEMD) in the monitoring of high-intensity focused ultrasound (HIFU) treatment is presented. Non-invasive surgeries such as HIFU have been developed for various therapeutic applications. In this technique, a suitable imaging method is necessary for monitoring of the treatment to achieve adequate treatment safety and efficacy. So far, several methods have been proposed, such as ultrasound radiofrequency (RF) signal processing techniques, and imaging methods such as X-ray, MRI, and ultrasound to monitor HIFU lesions. Results:In this paper, a CS-EMD method is used to detect the HIFU thermal lesion dimensions using different types of wavelet transform. The results of the processing on the real data demonstrate the potential for this technique in image-guided HIFU therapy. Conclusions: In this study, a new application of compressive sensing in the field of monitoring of the HIFU treatment is presented. To the best of our knowledge, so far no studies on compressive sensing have been carried out in the monitoring of the HIFU. Based on the results obtained, it was showed that the number of measurements and Intrinsic Mode Functions have the function of noise reduction. Moreover, it was shown that the successful reconstruction of the CS signals can be achieved using a threshold based algorithm. To this end, in this work it was shown that by selecting an appropriate number of measurements, the sparse transform, and a thresholding algorithm, we can achieve a more accurate detection of the HIFU thermal lesion size.
Keywords
Full Text:
PDFReferences
Clement GT. Perspectives in clinical uses of high-intensity focused ultrasound. Ultrasonics 2004;42:1087-93.
Lafon C, Inserm L, Bouchoux G, Souchon R, Chapelon JY. Monitoring and Follow up of HIFU Lesions by Ultrasound. In: 4th IEEE International Symposium on Biomedical Imaging: From Nano to Macro. Arlington 2007. p. 1068-71.
Zheng X, Vaezy S. An acoustic backscatter-based method for localization of lesions induced by high-intensity focused ultrasound. Ultrasound Med Biol 2010;36:610-22.
Tavakkoli J, Sanghvi NT. Ultrasound-guided HIFU and thermal ablation. In: Frenkel V, editor. Therapeutic Ultrasound: Mechanisms to Applications. Hauppauge, NY: Nova Science Publishers; 2011. p. 137-61.
Damianou C, Pavlou M, Velev O, Kyriakou K, Trimikliniotis M. High intensity focused ultrasound ablation of kidney guided by MRI. Ultrasound Med Biol 2004;30:397-404.
Khokhlova TD, Canney MS, Lee D, Marro KI, Crum LA, Khokhlova VA, et al. Magnetic resonance imaging of boiling induced by high intensity focused ultrasound. J Acoust Soc Am 2009;125:2420-31.
Zhang S, Wan M, Zhong H, Xu C, Liao Z, Liu H, et al. Dynamic changes of integrated backscatter, attenuation coefficient and bubble activities during high-intensity focused ultrasound (HIFU) treatment. Ultrasound Med Biol 2009;35:1828-44.
Rangraz P, Behnam H, Shakhssalim N, Tavakkoli J. A feed-forward neural network algorithm to detect thermal lesions induced by high intensity focused ultrasound in tissue. J Med Signals Sens 2012;2:192-202.
Coussios CC, Farny CH, Haar GT, Roy RA. Role of acoustic cavitation in the delivery and monitoring of cancer treatment by high-intensity focused ultrasound (HIFU). Int J Hyperthermia 2007;23:105-20.
Miller NR, Bamber JC, Meaney PM. Fundamental limitations of noninvasive temperature imaging by means of ultrasound echo strain estimation. Ultrasound Med Biol 2002;28:1319-33.
Chen H. Harmonic Motion Imaging in Abdominal Tumor Detection and HIFU Ablation Monitoring: A Feasibility Study in a Transgenic Mouse Model of Pancreatic Cancer. In: IEEE Ultrasonics Symposium. Chicago, IL, New York: IEEE; 2014. p. 923-6.
Adams M, Robin OC, Ronald AR. Treatment planning and strategies for acousto-optic guided highintensity focused ultrasound therapies. J Acoust Soc Am 2014;135:2267.
Alhamami M, Kolios MC, Tavakkoli J. Photoacoustic detection and optical spectroscopy of high-intensity focused ultrasound-induced thermal lesions in biologic tissue. Med Phys 2014;41:053502.
Huang NE, Shen Z, Long RS, Wu MC, Shih HH, Zheng Q, et al. The empirical mode decomposition and the Hilbert spectrum for nonlinear and non stationary time series analysis. Proc R Soc Lond 1998;25:903-95.
Balocchi R, Menicucci D, Santarcangelo E, Sebastiani L, Gemignani A, Ghelarducci B, et al. Deriving the respiratory sinus arrhythmia from the heartbeat time series using empirical mode decomposition. Chaos Solitons Fractals 2004;20:171-7.
Bouzid A, Ellouze N. Empirical mode decomposition of voiced speech signal. 2004First international symposium on control. Hammamet, Tunisia: Communications and Signal Processing 2004. p. 603-6.
Yu D, Cheng J, Yang Y. Application of EMD method and Hilbert spectrum to the fault diagnosis of roller bearings. Mech Syst Signal Process 2005;19:259-70.
Ghasemifard H, Behnam H, Tavakkoli J. Toward high-intensity focused ultrasound lesion quantification using compressive sensing theory. Proc Inst Mech Eng H 2017;231:1152-64.
Liebgott H, Prost R, Friboulet D. Pre-beamformed RF signal reconstruction in medical ultrasound using compressive sensing. Ultrasonics 2013;53:525-33.
Liebgott H, Basarab A, Kouame D. Compressive sensing in medical ultrasound. Lyon, France: IEEE Ultrasonics Symposium; 2012. p. 1-6.
Friboulet D, Liebgott H, Prost R. Compressive Sensing for Raw RF Signals Reconstruction in Ultrasound. Processing. San Diego, California, USA: IEEE Ultrasonics Symposium; 2010. p. 367-70.
Wan-Zheng N, Hai-Yan W, Kouame D, Bernard O, Friboulet D. The analysis of noise reduction performance in compressed sensing. Toulouse, France: IEEE International Conference Signal Processing, Communications and Computing (ICSPCC); 2011.
Donoho D. Compressed sensing. IEEE Trans Inform Ther 2006;52:1289-309.
Rahimian S, Tavakkoli J. Estimating dynamic changes of tissue attenuation coefficient during high-intensity focused ultrasound treatment. J Ther Ultrasound 2013;1:14.
Guangming SH, Danhua LI, Dahua GA, Liu DH, Gao DH, Liu Z, et al. Advances in theory and application of compressed sensing. Acta Electron Sin 2009;5:1070-81.
Donoho D, Johnstone I. Ideal spatial adaptation by wavelet shrinkage. Biometrika 1994;81:425-55.
Cobbold RS. Ultrasound Imaging Systems: Design, Properties and Applications, Foundations of Biomedical Ultrasound. New York: Oxford University Press; 2007. p. 512-3.
Dragomiretskiy K, Zosso D. Variational Mode Decomposition. IEEE Trans Signal Process 2014;62:531-44.
Refbacks
- There are currently no refbacks.
https://e-rasaneh.ir/Certificate/22728
ISSN : 2228-7477