Interpretation of In-air Output Ratio of Wedged Fields in Different Measurement Conditions

Parinaz Mehnati, Farideh Biglari, Ali Jomehzadeh

DOI: 10.4103/jmss.JMSS_36_18

Abstract


Background: The collimator scatter factor (Sc) is one of the most important parameters in monitor unit (MU) calculation. There are several factors that impact Sc values, including head structures, backscatter in dose monitoring chambers, and wedges. The objective of this study was to investigate the variation of Sc with different buildup cap materials, wall thickness of buildup caps, source-to-chamber distances (SCDs), ionization chambers, and wedge angles in 6 MV photon beam. Methods: In this study, copper and Perspex buildup caps were made with two different thicknesses for each buildup cap. Measurements were performed on an Elekta Compact medical linear accelerator (6 MV) using RK dosimeter with a sensitive volume of 0.120 cm3 and Farmer-type ion chamber with a sensitive volume of 0.65 cm3. In all measurements, buildup caps and ionization chambers were positioned such as to stand vertically to the beam central axis. It was also investigated the effect of internal wedge with different angles (30° and 60°) different SCDs on Sc. Results: It was found in large field sizes, Sc values in Perspex buildup cap were higher than copper. Different SCDs and type of ion chamber and wall thickness of buildup caps had no significant influence on Sc values. The presence of wedge influenced Sc values significantly. Variation of Sc in wedged fields compared to open fields had a maximum deviation of 0.9% and 6.8% in 30° and 60° wedge angles, respectively. Conclusion: It was found that the presence of wedges had a significant influence on Sc and increases with wedge angles. As such, it should be taken into account in manual MU calculations.


Keywords


Buildup cap; collimator scatter factor; ionization chamber; wedge

Full Text:

PDF

References


Jomehzadeh A, Shahbazi-Gahrouei D, Jahanbakhsh V. Effect of material and wall thickness buildup caps on the head scatter factor measurements in irregular fields shielded by cerrobend. J Med Signals Sens 2017;7:247-51.

Miyashita H, Hatanaka S, Fujita Y, Hashimoto S, Myojyoyama A, Saitoh H, et al. Quantitative analysis of in-air output ratio. J Radiat Res 2013;54:553-60.

Shahbazi-Gahrouei D, Changizi B, Jomehzadeh A, Larizadeh M.H. The effect of contrast media on treatment planning and dose calculation in radiation therapy of pelvis cancers. J Isfahan Med Sch 2017;34:1389-94.

Khan FM, Gibbons JP. Khan's the Physics of Radiation Therapy. 5th ed. Philadelphia: Lippincott Williams & Wilkins; 2014.

Appasamy M, Xavier S, Kuppusamy T, Velayudham R. Measurement of head scatter factor for linear accelerators using indigenously designed columnar mini phantom. Pol J Med Phys Eng 2011;17:13-25.

Iftikhar A. Measurements of output factors using different ionization chambers and build up caps. Iran J Radiat Res 2012;10:95-8.

Ashokkumar S, Nambiraj A, Sinha SN, Yadav G, Raman K, Bhushan M, et al. Measurement and comparison of head scatter factor for 7 MV unflattened (FFF) and 6 MV flattened photon beam using indigenously designed columnar mini phantom. Rep Pract Oncol Radiother 2015;20:170-80.

Li J, Zhu TC. Monte carlo simulation of the effect of miniphantom on in-air output ratio. Med Phys 2010;37:5228-37.

Birgani MJ, Chegeni N, Behrooz MA, Bagheri M, Danyaei A, Shamsi A, et al. An analytical method to calculate phantom scatter factor for photon beam accelerators. Electron Physician 2017;9:3523-8.

Zhu TC, Ahnesjö A, Lam KL, Li XA, Ma CM, Palta JR, et al. Report of AAPM therapy physics committee task group 74: In-air output ratio, Sc, for megavoltage photon beams. Med Phys 2009;36:5261-91.

Appasamy M, Xavier S, Kuppusamy T, Velayudham R. Measurement of the in-air output ratio for high-energy photon beams used in radiotherapy. Turk J Med Sci 2013;43:441-7.

Ashokkumar S, Nambi Raj NA, Sinha SN, Yadav G, Thiyagarajan R, Raman K, et al. Comparison of head scatter factor for 6MV and 10MV flattened (FB) and unflattened (FFF) photon beam using indigenously designed columnar mini phantom. J Med Phys 2014;39:184-91.

Ahmad M, Hussain A, Muhammad W, Rizvi SQ, Matiullah. Studying wedge factors and beam profiles for physical and enhanced dynamic wedges. J Med Phys 2010;35:33-41.

Appasamy M, Xavier SV, Kuppusamy T, Velayudham R. Study of head scatter factor in 4MV photon beam used in radiotherapy. Technol Cancer Res Treat 2013;12:207-16.

Chen L-X, Sun H-Q, Huang S-M, Sun W-Z, Gao X-W. Measurements and comparisons for data of small beams of linear accelerators. Chin J Cancer 2009;28:272-6.

Li J, Zhu TC. Measurement of in-air output ratios using different miniphantom materials. Phys Med Biol 2006;51:3819-34.

Chegeni N, Tahmasebi Birgani MJ, Hosseini SM, Einy F, Atarod M, Asgarian Dehkordy Z, et al. Measurement of collimator scatter factor for photon fields using gafchromic EBT2 film and ion chamber. Jundishapur Sci Med J 2016;14:641-52.

Warrener K, Hug B, Liu P, Ralston A, Ebert MA, McKenzie DR, et al. Small field in-air output factors: The role of miniphantom design and dosimeter type. Med Phys 2014;41:021723.

Jursinic PA. Measurement of head scatter factors of linear accelerators with columnar miniphantoms. Med Phys 2006;33:1720-8.


Refbacks

  • There are currently no refbacks.


 

  https://e-rasaneh.ir/Certificate/22728

https://e-rasaneh.ir/

ISSN : 2228-7477