Effect of Material and Wall Thickness Buildup Caps on the Head Scatter Factor Measurements in Irregular Fields Shielded by Cerrobend

Ali Jomehzadeh, Daryoush Shahbazi-Gahrouei, Vahid Jahanbakhsh

DOI: 10.4103/jmss.JMSS_10_17

Abstract


Background: The head scatter factor (Sc) is important to measurements radiation beam and beam
modeling of treatment planning systems used for advanced radiation therapy techniques. This study
aimed to investigate the design of a miniphantom to measurement variations in collimator Sc in the
presence of shielding blocks for shaping the beam using different field sizes. Materials and Methods:
Copper, Brass, and Perspex buildup caps were designed and fabricated locally as material with three
different thicknesses for buildup caps (miniphantoms). Measurements were performed on an Elekta
Compact medical linear accelerator (6 MV) in Shafa Kerman Hospital, Iran. The Farmer-type ion
chamber FG65-P (Scanditronix, Wellhofer) was used for all measurements. To measure the Sc,
miniphantom was positioned in a stand vertical to the beam central axis. Results: The data indicate
that the S
c measurements using different buildup cap materials and thicknesses in 5 × 10, 7.5 × 7.5, and
two 10 × 10 cm Cerrobend shield blocks ranged 0.98 to 1.00, 1.04 to 1.05, and 1.04 to 1.06, respectively.
Also, it was observed that by increasing the block shield area from 50 cm2 to both 56.25 and 100 cm2, the
S
c increased in all situations. Conclusion: Results showed that using Brass compared to Perspex and
Copper has less uncertainty due to its simple preparation and cutting which is useful to measurement of
variations in collimator S
c and shaping the photon beam.

Full Text:

PDF

References


Shahbazi-Gahrouei D, Gookizadeh A, Abdolahi M. Comparison of conventional radiotherapy techniques with different energies in treating prostate cancer, employing a designed pelvis phantom. J Med Sci 2008;8:429-32.

Rezaee V, Shahbazi-Gahrouei D, Monadi S, Saeb M. Evaluation of error doses of treatment planning software using solid anthropomorphic phantom. J Isfahan Med Sch 2016;34:908-13.

Li J, Zhu TC. Monte Carlo simulation of the effect of miniphantom on in-air output ratio. Med Phys 2010;37:5228-37.

Jursinic P. Measurement of head scatter factors of linear accelerators with columnar miniphantoms. Med Phys 2006;33: 1720-8.

Karbalee M, Shahbazi-Gahrouei D, Tavakoli MB. An approach in radiation therapy treatment planning: A fast GPU-based Monte Carlo method. J Med Signals Sens 2017;7:108-13.

Ishikura S. Quality assurance of radiotherapy in cancer treatment: Toward improvement of patient safety and quality of care. Jpn J Clin Oncol 2008;38:723-9.

Sigamani A, Arunai N, Sujit Nath S, Girigesh Y, Rajesh T, Kothanda R, et al. Comparison of head scatter factor for 6 MV and 10 MV flattened (FB) and unflattened (FFF) photon beam using indigenously designed columnar miniphantom. J Med Phys 2014;39:184-91.

Tahmasbi-Birgani M, Seif F, Bayatiani MR. Dosimetric characteristics of mercury and cerrobend blocks in megavoltage radiation therapy. J Radioanal Nucl Chem 2015;303:1843-50.

Zhu T, Ahnesjo A, Lam KL, Li XA, Ma CM, Palta JR. Report of AAPM Therapy Physics Committee Task Group 74: In-air output ratio, Sc, for megavoltage photon beams. Med Phys 2009;36: 5261-91.

Eaton D, Twyman N, Thomas S. Commissioning a miniature multileaf collimator for small field radiotherapy. Med Dosim 2010;35:1-6.

Farajollahi A, Bouzarjomehri F, Kiani M. Comparison between clinically used irregular fields shielded by Cerrobend and standard lead blocks. J Biomed Phys Eng 2015;5:77-82.

Iftikhar A, Wazir M, Kakakhail MB, Sbilal A, Amjad H, Khwaja A. Comparison of lead and Cerrobend blocks for incident photon flux of 6 and 15 MV X-rays. Iran J Cancer 2011;10:10-4.

Iftikhar A. Measurements of output factors using different ionization chambers and build up caps. Iran J Radiat Res 2012;10:10-4.

Mukhtar A, Samat SB, Dolah MT. Long term stability of farmer type ionization chamber calibration coefficient belonging to local radiotherapy centers in Malaysia. Malays J Anal Sci 2014;18:7-14.


Refbacks

  • There are currently no refbacks.


 

  https://e-rasaneh.ir/Certificate/22728

https://e-rasaneh.ir/

ISSN : 2228-7477